OxiTop ® -C Measuring Head
Instead of the display and keys, the OxiTop ® -C measuring head has an infrared interface with which it communicates with Controller OC 100 or OC 110. By "e;pointing"e; the controller at an OxiTop ® -C measuring head it can be identified and started, data can be called up or deleted and sampling progress can be displayed.Each measuring head has its own identification number; this means that manual identification of samples is…
The supplier does not provide quotations for this product through SelectScience. You can search for similar products in our Product Directory.
Instead of the display and keys, the OxiTop ® -C measuring head has an infrared interface with which it communicates with Controller OC 100 or OC 110. By "e;pointing"e; the controller at an OxiTop ® -C measuring head it can be identified and started, data can be called up or deleted and sampling progress can be displayed.
Each measuring head has its own identification number; this means that manual identification of samples is no longer necessary, even for parallel samples. In addition, statistical evaluations can be easily performed for parallel samples.
The OxiTop ® -C measuring heads have an AutoTemp function for delaying the start of samples which are too cold by up to 4 hours. This mode can be deactivated for BOD standard.
The measuring heads can store up to 360 data sets. Data are automatically stored in the corresponding interval according to the interval period set (0.5 h to 99 days).
The built-in pressure sensor can register differences in pressure ranging from 500 to 1,350 hPa.