Electronically Controlled Load: Automated Self-Diagnostic Check at the Touch of a Button
13 Nov 2018The operator simply holds the mode key to first perform a self-diagnostic check and proceed to ECL where internally a repeatability test is performed without use of an external weight. This critically checks the stability of analog-to-digital conversion before applying a controlled load inside the balance (between 0.3% to 3% of the weighing capacity). It does this by altering the equilibrium state of the weigh sensor and takes 10 repeated measurements. The entire process takes only 1.5 minutes versus traditional repeatability procedures which can take up to 10 minutes, ultimately giving greater convenience and time savings to customers. The ECL technology provides valuable information which Auto Precision Assessment (APA) smartly processes into meaningful, beneficial outcomes such as Dia-Check, Standard Deviation, Quick Min-S, and Min-S-Alert.