MED64 Multi-electrode Array System

Bandwidth of 0.1Hz -10 kHz allows users to record several types of extracellular potentials. • Low-cut filter (High pass filter) with 0.1, 1, 10, and 100Hz cutoff. • High cut filter (Low pass filter) with 1, 2.5, 5, 7.5, and 10kHz cutoff. • Built-in 2-channel stimulator allows users to program and deliver complex stimulation sequences independently through any 2 electrodes. • USB connection to computer.SpecificationsHead Ampl…

AutoMate Scientific Inc.

The supplier does not provide quotations for this product through SelectScience. You can search for similar products in our  Product Directory.

Ease of Use
After Sales Service
Value for Money
Write your own review

Bandwidth of 0.1Hz -10 kHz allows users to record several types of extracellular potentials.
• Low-cut filter (High pass filter) with 0.1, 1, 10, and 100Hz cutoff.
• High cut filter (Low pass filter) with 1, 2.5, 5, 7.5, and 10kHz cutoff.
• Built-in 2-channel stimulator allows users to program and deliver complex stimulation sequences independently through any 2 electrodes.
• USB connection to computer.

Specifications

Head Amplifier
Number of channels 64
Gain x 10
Bandwidth 0.1Hz-100 kHz (+0dB to -3dB)
Input impedance 100MΩ
Output impedance 10kΩ

Stimulator
Number of channels 2
Output format: Constant current
Maximum input voltage +/-4V
Maximum output current +/-200µA

---

Main Amplifier
Number of channels 64
Gain 1-217
Bandwidth 0.1Hz-10 kHz (+0dB to -3dB)
Analog high cut filter 1, 2.5, 5, 7.5, 10kHz (-6dB/oct)
Analog low cut filter 0.1, 1, 10, 100Hz (-12dB/oct)
Input impedance 100 MΩ

Digitizer
Resolution 16 bit
Sampling rate 20kHz
Output: USB

Product Overview