Objective Heater Controller
To eliminate the heat loss from the specimen, Bioptechs developed a patented Objective Heater System®, which includes a heater/sensor and an electronic controller specifically designed for this purpose. The heater/sensor is comprised of an adjustable thin-film heating band which surrounds 3/4 of the diameter of the upper region of the central retracting tube of the objective. A temperature probe positioned in the gap formed be…
The supplier does not provide quotations for this product through SelectScience. You can search for similar products in our Product Directory.
To eliminate the heat loss from the specimen, Bioptechs developed a patented Objective Heater System®, which includes a heater/sensor and an electronic controller specifically designed for this purpose. The heater/sensor is comprised of an adjustable thin-film heating band which surrounds 3/4 of the diameter of the upper region of the central retracting tube of the objective. A temperature probe positioned in the gap formed between the ends of the heating band provides accurate feedback to a closed loop controller. A metal cube shaped frame supports a thermal sensor and contains a mechanism to adjust the size of the heater-band.
The Objective Controller® is specifically designed to slowly heat the objective over a fifteen minute warm-up period, then hold the objective at the setpoint value within 0.1°C. The Controller operates from ambient to 50°C and has special safety circuitry which utilizes a 0.9°C error window to shut down the Controller and sound an alarm if the temperature of the objective deviates after it has reached setpoint.